CONDUCTIVITY

C1
  • ceramic cell 

  • inside diameter: 16 mm

  • Ti protection housing

  • length: 115 mm

  • bulkhead terminal

main applications
  • calibration reference
  • high precision deep sea research
  • long term monitoring
  • extreme high temperature environments ( hot vents)
  • fresh water measurements with low range and the corresponding better resolution
principle symmetric 7-pole-cell

sensitivity strongly restricted to the cell interior

range 0...65 mS/cm
resolution 0.001 mS/cm
accuracy 0.01 mS/cm
time const. depending on water exchange time
available interfaces
  • 16 to 18 bit serial synchronous or asynchronous

 

C2
  • ceramic cell

  •  inside diameter: 8 mm

  • Ti protection housing

  • length: 43 mm

  • bulkhead terminal

main applications
  • high precision deep sea research
  • oceanographic profiling
  • extreme high temperature environments ( hot vents)
  • fresh water measurements with low range and the corresponding better resolution
principle symmetric 7-pole-cell

sensitivity strongly restricted to the cell interior

range 0...65 mS/cm
resolution 0.001 mS/cm
accuracy 0.02 mS/cm
time const. depending on water exchange time
available interfaces
  • analog voltage or current 
  • 16 bit serial asynchronous

 

C3

 
  • ceramic cell inside diameter: 16 mm

  • length: 115 mm

  • bulkhead terminal

main applications
  • long term monitoring down to 300 m
principle symmetric 7-pole-cell

sensitivity strongly restricted to the cell interior

range 0...65 mS/cm
resolution 0.02 mS/cm
accuracy 0.05 mS/cm
time const. depending on water exchange time
available interfaces
  • analog voltage or current 
  • 16 bit serial asynchronous

 

C4
  • ceramic cell inside diameter: 8 mm

  • moulded plastic housing

  • length: 43 mm

  • bulkhead terminal

main applications
  • profiling down to 300 m
  • fresh water measurements with low range and the corresponding better resolution
principle symmetric 7-pole-cell

sensitivity strongly restricted to the cell interior

range 0...65 mS/cm
resolution 0.02 mS/cm
accuracy 0.05 mS/cm
time const. depending on water exchange time
available interfaces
  • analog voltage or current 
  • 16 bit serial asynchronous